Model |
MS3006 |
Measurement Geometry |
6 measurement angles (6 illumination sources, 1 receivers) |
Measure Angle |
45° Receiver: 45as-15°, 45as15°, 45as25°, 45as45°, 45as75°, 45as110° Conform to the standards: ASTM D 2244, E 308, E 1164, E 2194, E2539, DIN 5033, 5036, 6174, 6175-1,6175-2; ISO 7724, 11664-4 SAE J 1545 |
Application |
Provide accurate and consistent color measurement for metallic, pearlescent and other complex special effect color products |
Light Source |
Full spectrum LED light source with blue enhancement |
Lamp Life |
5 years, 3 million times measurements |
Spectrophotometric Mode |
Concave Grating |
Sensor |
256 Image Element Double Array CMOS Image Sensor |
Wavelength Range |
400nm-700nm |
Wavelength Interval |
10nm |
Measurement Range |
0~600% |
Semiband Width |
10nm |
Measuring Aperture |
Illumination size Φ23mm / Measuring spot 9X12mm (Customized: Illumination size Φ10mm / Measuring spot 6X8mm) |
Color Space |
CIE LAB, XYZ, Yxy, LCh, βxy, DIN Lab99 |
Color Difference Formula |
ΔE*ab,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE DIN6175 |
Other Colorimetric Index |
Flop Index, Int-Em |
Observer Angle |
2°/10° |
Illuminant |
D65, A, C, D50, D55, D75, F1, F2(CWF), F3, F4, F5, F6, F7(DLF), F8, F9, F10(TPL5), F11(TL84), F12(TL83/U30) |
Display |
Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Offset, Color simulation, Sample effect value, Effect difference value |
Measuring Time |
Approx. 1 second for one angle Approx. 6 seconds for all angles |
Repeatability |
Spectral reflectance: Standard deviation within 0.08% Chromaticity value: ΔE*ab 0.03 ( When a white calibration plate is measured 30 times at 5 second intervals after white calibration) |
Reproducibility |
ΔE*<0.10,avg on the gray tile of BCRA tile set ΔE*<0.25,avg on the color BCRA tile set |
Inter-Instrument Error |
0.2ΔE*00(avg on reference Series II BCRA tile set) |
Effect Parameters |
Sparkle Grade(SG),Diffuse coarseness(DC) and Color Variation(CV) |
Effect Measurement |
6 angles Sparkle Grade(SG),Color Variation(CV):15as-45°,15as-30°,15as-15°,15as15°,15as45°,15as80° 15d Diffuse coarseness(DC) |
Effect Repeatability |
Sparkle Grade(SG) Short-term repeatability: 0.12% (10 times standard deviations) (When a color plate is measured 10 times at 10 second intervals after white calibration) Diffuse coarseness(DC) Short-term repeatability:e0.09% (10 times standard deviations) (When a color plate is measured 10 times at 10 second intervals after white calibration) |
Effect Reproducibility |
Sparkle Grade(SG) Reproducibility: 1.9% (10 times standard deviations) (avg on reference Series II BCRA tile set) Diffuse coarseness(DC) Reproducibility: 1.4% (10 times standard deviations)(avg on reference Series II BCRA tile set) |
Trigger mode |
Pressure sensing trigger, button trigger, software trigger |
Measuring Mode |
Single measurement, average measurement (1-99), continuous measurement (1-99) |
Locating Mode |
Color camera preview |
Dimension |
L*W*H=195X83X128mm |
Weight |
About 1Kg |
Power |
lithium-ion battery, 3.7V,5000mAh, Continuous test 6000 times within 8 hours of full charge |
Display |
3.5-inch TFT color LCD, Capacitive Touch Screen |
Interface |
USB, Bluetooth 5.0 |
Data Storage |
1000 pcs Standards, 4000 pcs Samples |
Language |
Simplified Chinese, Traditional Chinese, English |
Operating Environment |
10℃ to 50℃, humidity does not exceed 85%, no condensation |
Storage Environment |
-20℃ to 50℃, humidity does not exceed 85%, no condensation |
Calibration |
Built-in white board parameters, external white board, black cavity |
Calibration Interval |
4 hours,8 hours,24 hours,Startup calibration |
Standard accessories |
Power Adapter, USB Cable, User Guide,PC Software(download from the official website), Calibration Board, black light trap,Protective cap, wristband |
Optional Accessory |
Micro-printer |
Notes |
Technical parameters are for reference only, subject to actual sales products. |