Hot Wire versus Vane Anemometers

Sper Scientific Anemometer

Anemometers measure air velocity. There are two basic types for different applications:

Hot Wire Anemometer

Thermoelectric ("Hot Wire") Anemometers

Heat a wire to a specified temperature and then measure the rate of cooling.
This rate is proportional to air speed. Thermoelectric measurement provides fast response times and excellent sensitivity to very low air flows of <1 m/s, but these probes are also the most delicate of the two types and are not suitable for environments which are dusty, humid, corrosive or where there are rapid fluctuations in the ambient temperature, all of which effect the rate of cooling. These anemometers cover ranges of 0.01 m/s ~ 5 m/s.

 

Vane Anemometer

Vane Anemometers

Work on the principle that a freely turning turbine will rotate at a speed directly proportional to the wind speed. Vane Anemometers cover ranges as low as 0.15 m/s and as high as 40 m/s, or even higher by special order. Useful for a wide variety of applications including relatively harsh environments. Thresholds are not as low as with thermoelectric anemometers since the vane must overcome initial friction before it will turn.

See the full line of Sper Anemometers here.



Publicación más antigua Publicación más reciente